The analyzer was tested by using Pierce & Extractor gun only, as suggested by Demers.
1. Dynamic range of the analyzer
The analyzer detector layout is shown as below: (If seen from the backside)
The ion beam energy = 10keV
----------------------------------------------------------------------------------------------
50% on T1, 50% on T3, Va=2.865kV
50% on C1, 50% on C3, Va=3.375kV
50% on B1, 50% on B3, Va=4.105kV
50% on T1, 50% lost above T1, Va=2.735kV
50% on T3, 50% lost between T3 & C1, Va=3.025kV
50% on C, 50% lost between T3 & C1, Va=3.2kV
50% on C3, 50% lost between C3 & B1, Va=3.59kV
50% on B1, 50% lost between C3 & B1, Va=3.82kV
50% on B3, 50% lost below B3, Va=4.41kV
----------------------------------------------------------------------------------------------
2. Gain curve
a) With center slit, center detectors: Wb=10keV
Angle | 23 | 24 | 25 | 26 | 27 | 28 | 29 | 30 | 31 | 32 | 33 | 34 | 35 | 36 |
Va(kV) | 3.42 | 3.41 | 3.39 | 3.385 | 3.375 | 3.375 | 3.375 | 3.375 | 3.375 | 3.375 | 3.375 | 3.37 | 3.365 | 3.36 |
b) With center slit, top detectors: Wb=10keV
Angle | 23 | 24 | 25 | 26 | 27 | 28 | 29 | 30 | 31 | 32 |
Va(kV) | 2.69 | 2.71 | 2.745 | 2.78 | 2.815 | 2.85 | 2.85 | 2.875 | 2.89 | 2.91 |
c) With center slit, bottom detector: Wb=10keV
Angle | 24 | 25 | 26 | 27 | 28 | 29 | 30 | 31 | 32 | 33 | 34 | 35 |
Va(kV) | 4.52 | 4.4 | 4.315 | 4.26 | 4.2 | 4.175 | 4.115 | 4.06 | 4.01 | 3.975 | 3.965 | 3.925 |
3. Calibration of the 2nd set toroidal sweep plates in the secondary beamline
By using the analyzer detectors c1 & c2, Wb=10keV
Conditions | Voltage on the plates |
Just to the left of C1 | 201.6V |
50% on C1, 50% to the left of C1 | 174V |
Fully on C1 | 12.4 ~ 150V |
50% on C1, 50% on C2 | -22.4V |
Fully on C2 | -45.2~-140V |
50% on C2, 50% lost to the right of C2 | 193.2V |
Just to the right of C2 | 209.6V |